Skip to main content

SE706 Oxygen Sensor | Memosens | 220 mm | Ex | Measuring range (resolution) 0 ... 50 mg/l (6 μg/l)

Description

SE706 Oxygen Sensor | Memosens | 220 mm | Ex | Measuring range (resolution) 0 ... 50 mg/l (6 μg/l)

Sensor in robust and hygienic stainless steel design, high accuracy and low detection limit, simple and fast maintenance due to special membrane, digital, with Memosens technology.

  • perfect galvanic isolation thanks to Memosens technology
  • no influence of humidity in the connector
  • precalibration in the lab
  • digital data transfer
  • integrated sensor diagnostics
  • high level of process safety
  • durable materials
  • robust design
  • easy to clean thanks to extremely smooth surface
  • cost-efficient service due to easy-to-replace membrane module and interior body
  • hygienic design
  • high resolution of 6 ppb
  • operation also in hazardous locations

Reliable in-line measurement of dissolved oxygen, even under sterile and hygienic conditions. High signal stability. Simple maintenance due to modular design.

Special membrane with steel mesh and PTFE coating. Sterilizable, autoclavable and CIP capable. Also suitable for use in hazardous locations.

Applications
Biotechnology, pharmaceutical industry, fermentation, various fields of analytical chemistry

We look forward to hearing from you!
Contact Us

Technical details

Measurement Parameter: Oxygen
Permissible pressure range (measurement): 0.2 ... 6 bar absolute
Connector: Memosens
Explosion protection: Ex
Length: 220 mm
Measuring Range: pO₂ < 1200 mbar
Temperature Detector: NTC 22 kΩ
Measuring Principle: Amperometric
Membrane material: PTFE/silicone/PTFE, FDA compliant (reinforced with steel mesh)
Flow dependence: ≤ 5 %
Signal Transmission: digital
Mechanical pressure resistance: Max. 12 bar absolute
Temperature range (stability): –5 ... +135 °C (steam sterilizable/autoclavable)
Accuracy: 1 % + 4 ppb
Signal in ambient air: 50 ... 110 nA
Response time at 25 °C (air –> N2): 98 % full scale < 90 s
Residual signal in O₂-free medium: ≤ 0.1 % of signal in ambient air
O-ring material (process-wetted): silicone (FDA & USP Class VI tested)
Process-wetted sensor parts: Stainless steel 1.4404 (material certificate 3.1B)
Surface roughness: N 5 (RA < 0.4 μm)
Interior body incl. electrodes: replaceable (ZU 0567)

Downloads

Manuals

[16286] User Manual (DE, EN, KO)
Memosens Cable CA/MS
Version 09
[3061] User Manual
SE706 Memosens Oxygen Sensor
Version 04
[17421] User Manual
SE706 Memosens
incl. Ex Approval Japan (JPEx DEK19.0046X)
Version 04

Brochures

[13395] Brochure
Sensors - For measuring pH/ORP, conductivity or oxygen. Digital or analog.

Data Sheets

[3052] Data sheet
SE706 Memosens Oxygen Sensor
Product Description | Specifications | Product Range
[10145] Data sheet
Accessories for Sensors for Oxygen Measurement
Product Description | Specifications | Product Range

Certificates

[16639] Type Approval Certificate for Ex-Equipment (JA)
No. DEK19.0046X, Issue No. 0 (Japan)
SE5**X/*-*MSN | SE7**X/*-*MSN
[11064] EU Declaration of Conformity
No. EU210427A | 2021-04-27
Memosens Sensors | Types SE 5..X..-.MS., SE604X-MS**, SE605….., SE6**X**MS, SE7**…
[12902] Confirmation / Bestätigung
Series SE706X/*-NMSN, SE707X/*-NMSN
No components of animal origin / Kein tierischer Ursprung
[14685] Certificate of Conformity (FM)
No. FM17US0208X | 2019-05-27 | USA
Memosens Sensors | Types SE 5**, SE 6** and SE 7**
[16292] Korean safety certificate Kosha (KO)
No. 2020-KA4BO-0368X
Measuring Cable CA/MS
[12826] EU-Type Examination Certificate (DEKRA)
Nr. BVS 16 ATEX E 037 X | 2024-02-20
Memosens-Ex-Sensors
[14155] Declaration of Compliance 2.1 (DE, EN)
according to EN 10204 | gemäß EN 10204

Control Drawings

[14185] Control Drawing
Inductive Sensor-Cable Connection System Memosens
Zeichnungsnummer 214.410-240

Our service

Used in Process or Applications

The taste and quality of beer are highly dependent on the quality of the water used for brewing it. It is determined by pH and conductivity measurements.

Related Products

Relevant Blog Posts

Knick Products Comply With UL - What Does That Mean?

There are different assessments and approaches to reach an adequate level of safety. Manufacturers are typically on the “safe” side, if they comply with recognized standards.