Passer au contenu principal

703 Laboratory Conductivity Meter | Automatic device test (Fullcheck) and calibration | Records for QM documentation according to ISO 9000 and GLP

Description

703 Laboratory Conductivity Meter | Automatic device test (Fullcheck) and calibration | Records for QM documentation according to ISO 9000 and GLP

Quality assurance and measurement documentation in accordance with GLP

  • Measuring ranges from 0.000 µS/cm to 2000 mS/cm
  • 1.000 µS/cm … >1000 mS/cm with one sensor
  • Records for QM documentation according to ISO 9000 and GLP
  • Calibrated analog recorder output, galvanically isolated
  • Sensoface: monitoring of the sensor and equipment
  • Automatic calibration with standard solutions
  • EMC according to NAMUR
  • RS-232 interface for computer and printer
  • Simultaneous display of two measurement values
  • Self-contained clock
  • Leak-proof membrane keyboard
  • Robust housing
  • Protection category IP 54
  • 3-year warranty

The requirements for lab measurements become stricter every day. Quality assurance and measurement documentation in accordance with GLP are a must in many areas.

Fullcheck
automatically checks the device functions during power-on. Also during operation, a complete instrument check can be carried out at a single keystroke. Here, also display and keypad are checked besides the electrical characteristics.

Record printouts
With record printouts of the device self-test, the calibration, and the parameter settings, it is possible (as part of quality management to ISO 9000 and GLP) to document the serviceability and the regular maintenance and calibration of the unit.

Sensoface
checks the electrode and provides information on the electrode condition. The zero, slope, response time, and glass impedance of the electrode as well as the calibration interval are evaluated.

 

Nous attendons avec impatience votre réponse!
Nous Contacter

Détails techniques

Boîtier: Polyamide 12 renforcé de verre, couvercle en acier inox, protection IP 54, prêt à être raccordé au support ZU 6954

Téléchargements

Manuals

78598_KNF_703_s-2152564

Fiches techniques

[781] Data sheet
703 Laboratory Conductivity Meter
Product Description | Specifications | Product Range